A Confused Pacemaker
Loading...
Files
Date
Journal Title
Journal ISSN
Volume Title
Publisher
W.B. Saunders
Abstract
The electrocardiogram (ECG) raises the question of inappropriate device behavior and the possibility of ventricular lead oversensing causing failure of ventricular pacing. Careful analysis of the ECG proved that the mode of pacing was set to managed ventricular pacing (MVP) mode. The MVP mode should not be used in the setting of a complete atrioventricular conduction block. © 2016 Elsevier Inc.
Description
Keywords
Cardiac, Electrocardiogram, Heart, Pacemaker, Adult, Atrioventricular block, Electrocardiography, Female, Humans, Pacemaker, artificial, Aorta valve replacement, Article, Artificial heart pacemaker, Atrioventricular conduction, Complete heart block, Follow up, Heart ventricle pacing, Human, Mitral valve replacement, Priority journal, Rheumatic heart disease, Adverse effects, Case report, Pathophysiology