Structural valve deterioration of the Labcor Dokimos aortic prosthesis A single centre experience

dc.contributor.authorSawaya, Fadi
dc.contributor.authorIskandarani, Dounia
dc.contributor.authorChaabo, Omar
dc.contributor.authorGharzeddine, Walid
dc.contributor.authorSfeir, Pierre
dc.contributor.authorObeid, Mounir
dc.contributor.authorGhazzal, Ziyad
dc.contributor.authorRebeiz, Abdallah
dc.date.accessioned2023-08-09T08:25:40Z
dc.date.available2023-08-09T08:25:40Z
dc.date.issued2022
dc.description.abstractAim: The goal of this study was to assess the performance and incidence of the deterioration of the Labcor Dokimos bioprosthetic aortic valve. Methods: We performed a retrospective medical chart review of 116 patients who underwent surgical aortic valve replacement with the Labcor Dokimos aortic valve between 2010 and 2018. Abstracted data included patient demographic and echocardiographic data. Patients were divided into 2 groups: patients with structural valve deterioration (SVD) and patients without SVD. Results: Among the patients with complete follow-up (n=95), 10 patients were excluded because they died within a year; 85 patients were included in the final analysis. Of the 85 patients, 32 (38%) developed SVD; 22 (26%) had severe SVD, 15 (18%) of whom underwent reintervention. The most common etiology of SVD was severe central aortic regurgitation, which was detected in 91% of the patients who had severe SVD. The average time from operation to severe SVD was 4.7 years with a minimum of 1.5 years and a maximum of 7.9 years. Conclusion: Bioprosthetic aortic valve deterioration due to severe aortic regurgitation is common and occurs early with the Labcor Dokimos valve. This occurrence needs to be furthered investigated in larger registries.en_US
dc.identifier.urihttp://hdl.handle.net/10938/24111
dc.language.isoenen_US
dc.subjectvalve diseaseen_US
dc.subjecteducation-cardiacen_US
dc.subjecttranscatheter valve therapyen_US
dc.titleStructural valve deterioration of the Labcor Dokimos aortic prosthesis A single centre experienceen_US
dc.typeArticleen_US

Files

Original bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Sawaya_Structural valve deterioration_2021.pdf
Size:
356.59 KB
Format:
Adobe Portable Document Format
Description:
main article, 2022

License bundle

Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
license.txt
Size:
1.65 KB
Format:
Item-specific license agreed upon to submission
Description:

Collections