Experimental Insight into the Hemodynamics and Perfusion of Radiological Contrast in Patent and Non-patent Aortic Dissection Models

dc.contributor.authorSalameh, Elie
dc.contributor.authorSaade, Charbel
dc.contributor.authorOweis, Ghanem F.
dc.contributor.departmentDepartment of Mechanical Engineering
dc.contributor.departmentDivision of Health Professions
dc.contributor.departmentMedical Imaging Sciences
dc.contributor.facultyMaroun Semaan Faculty of Engineering and Architecture (MSFEA)
dc.contributor.facultyFaculty of Health Sciences (FHS)
dc.contributor.institutionAmerican University of Beirut
dc.date.accessioned2025-01-24T11:32:35Z
dc.date.available2025-01-24T11:32:35Z
dc.date.issued2019
dc.description.abstractPurpose: In a curved vessel such as the aortic arch, the velocity profile closer to the aortic root is normally skewed towards the inner curvature wall, while further downstream along the curve, the velocity profile becomes skewed towards the outer wall. In an aortic dissection (AD) disease, blood velocities in the true lumen (TL) and false lumen (FL) are hypothesized to depend on the proximity of the entry tear to the root of aortic arch. Faster velocity in the FL can lead to higher hemodynamic loading, and pose tearing risk. Furthermore, the luminal velocities control the perfusion rate of radiological contrast media during diagnostic imaging. The objective in this study is to investigate the effect of AD disease morphology and configuration on the blood velocity field in the TL and FL, and on the relative perfusion of radiological enhancement agents through the dissection. Methods: Eight in vitro models were studied, including patent and non-patent FL configurations. Particle image velocimetry (PIV) was used to quantify the AD velocity field, while laser-induced fluorescence (LIF) was implemented to visualize dynamical flow phenomena and to quantify the perfusion of injected dye, in mimicry of contrast-enhanced computed tomography (CT). Results: The location of the proximal entry tear along the aortic arch in a patent FL had a dramatic impact on whether the blood velocity was higher in the TL or FL. The luminal velocities were dependent on the entry/reentry tear size combination, with the smaller tear (whether distal or proximal) setting the upper limit on the maximal flow velocity in the FL. Upon merging near the distal reentry tear, the TL/FL velocity differential gave rise to the roll up and shedding of shear layer vortices that convected downstream in close proximity to the wall of the non-dissected aorta. In a non-patent FL, the flow velocity was practically null with all the blood passing through the TL. LIF imaging showed much slower perfusion of contrast dye in the FL compared to the TL. In a patent FL, however, dye had a comparable perfusion rate appearing around the same time as in the TL. Conclusions: Blood velocities in the TL and FL were highly sensitive to the exact dissection configuration. Geometric case A1R, which had its proximal entry tear located further downstream along the aortic arch, and had its entry and reentry tears sufficiently sized, exhibited the highest FL flow velocity among the tested models, and it was also higher than in the TL, which suggest that this configuration had elevated hemodynamic loading and risk for tearing. In contrast-enhanced diagnostic imaging, a time-delayed acquisition protocol is recommended to improve the detection of suspected cases with a non-patent FL. © 2019, Biomedical Engineering Society.
dc.identifier.doihttps://doi.org/10.1007/s13239-019-00407-4
dc.identifier.eid2-s2.0-85066063269
dc.identifier.pmid30805874
dc.identifier.urihttp://hdl.handle.net/10938/27833
dc.language.isoen
dc.publisherSpringer New York LLC
dc.relation.ispartofCardiovascular Engineering and Technology
dc.sourceScopus
dc.subjectContrast computed x-ray tomography (ct)
dc.subjectLif
dc.subjectMagnetic resonance (mr)
dc.subjectPatent false lumen
dc.subjectPiv
dc.subjectShear layer vortex
dc.subjectStanford type a
dc.subjectTear propagation
dc.subjectAneurysm, dissecting
dc.subjectAorta, thoracic
dc.subjectAortic aneurysm
dc.subjectBlood flow velocity
dc.subjectContrast media
dc.subjectHemodynamics
dc.subjectHumans
dc.subjectInjections
dc.subjectModels, cardiovascular
dc.subjectRegional blood flow
dc.subjectRheology
dc.subjectVascular patency
dc.subjectArches
dc.subjectBlood
dc.subjectBlood vessels
dc.subjectComputerized tomography
dc.subjectDiagnosis
dc.subjectFlow velocity
dc.subjectImage enhancement
dc.subjectMagnetic resonance
dc.subjectPatents and inventions
dc.subjectVelocity
dc.subjectVelocity measurement
dc.subjectVortex flow
dc.subjectContrast medium
dc.subjectComputed x-ray tomographies
dc.subjectShear-layer vortices
dc.subjectStanford
dc.subjectAortic arch
dc.subjectAortic dissection
dc.subjectArticle
dc.subjectComputational fluid dynamics
dc.subjectComputer assisted tomography
dc.subjectComputer simulation
dc.subjectDiagnostic imaging
dc.subjectFlow rate
dc.subjectFluorescence imaging
dc.subjectHeart cycle
dc.subjectHuman
dc.subjectLaser induced fluorescence spectroscopy
dc.subjectParticle image velocimetry
dc.subjectPerfusion
dc.subjectPriority journal
dc.subjectRefraction index
dc.subjectSimulation
dc.subjectVideorecording
dc.subjectX-ray computed tomography
dc.subjectBiological model
dc.subjectBlood flow
dc.subjectComparative study
dc.subjectDissecting aneurysm
dc.subjectFlow kinetics
dc.subjectInjection
dc.subjectPathophysiology
dc.subjectThoracic aorta
dc.subjectShear flow
dc.titleExperimental Insight into the Hemodynamics and Perfusion of Radiological Contrast in Patent and Non-patent Aortic Dissection Models
dc.typeArticle